+1 541-471-7111Contact Sales
Serinus 30 Trace

Serinus® 30 Trace Carbon Monoxide Analyzer

Exceptional trace CO precision for reliable baseline air monitoring

The Serinus® 30 Trace Carbon Monoxide (CO) Analyzer delivers precise and reliable performance at excellent value. It uses proven Non-Dispersive Infrared (NDIR) gas filter correlation technology combined with specialized trace-level hardware to measure low-level CO in ambient air. This allows the Serinus® 30T to sample in the most useful range for CO background measurement, representing the range of global CO mixing ratios in the unpolluted atmosphere. (LDL 20 ppb, range 0 to 20 ppm / 20,000 ppb).

Serinus® 30 Trace is US EPA approved (RFCA–0509–174).

Trace-Level Enhancements:

To achieve exceptional sensitivity and a lower detectable limit of just 20 ppb, the Serinus® 30T features several hardware upgrades over the standard Serinus® 30:

  • High-Resolution Measurement Cell: Features extra passes in the measurement cell to increase optical resolution.
  • Advanced IR Detector: Utilizes a unique, more sensitive IR detector with a larger area.
  • Active Cooling: The detector is housed in a thermally stabilized enclosure operating at a lower temperature (-50 °C) to significantly reduce thermal noise.
  • Sample Conditioning: Includes a sample dryer to remove water vapor interference (also available without a dryer as an option).

Platform Benefits:

Serinus® analyzers have been designed using our experience and knowledge gained from operating large air quality monitoring networks for more than 40 years. The result is instruments that integrate seamlessly into continuous monitoring networks:

  • Reliable performance complemented with a two-year warranty.
  • Minimize time spent on site performing maintenance through superior remote instrument control, diagnostic viewing, and calibration.
  • Comprehensive data logging and remote viewing of over 200 operational parameters.
  • Programming, viewing, downloading, and emailing of data is made even simpler using optional Bluetooth connectivity combined with the Serinus® Remote Android App.
  • Enhanced operator safety through the use of 12 VDC internal voltages.
  • Reduced cost of spare parts, accessories, and consumable items through extensive use of standard components across the Serinus® gas analyzer range.
  • Rack slide design makes accessing internal components and removing the analyzer from a rack cabinet easy.
  • AQMS power usage can be reduced due to the instrument’s low power demand and its ability to operate over a wide temperature range.

Designations

US EPA (RFCA–0509–174).

Operations / Interface

Communication:

  • #1 RS232 port (supports all protocols*)
  • #2 RS232 port (supports all protocols* and multidrop)
  • USB port (Advanced protocol only)
  • Bluetooth (Advanced protocol only)
  • TCP/IP network port (optional, supports all protocols*).

* Available protocols: Modbus (RTU/TCP), Bayern Hessen, EC9800 or Advanced

Communication Analog and Digital I/O Port:

  • 3 x analog outputs (current or voltage), 12 bit resolution user definable range (between two user specified points)
    • Current range: 0 – 20 mA, 2 – 20 mA or 4 – 20 mA
    • Voltage range: 0 – 1V, 0 – 5 V, 0 – 10 V
    • Voltage offset of 0 V, 0.25 V or 0.5 V
  • 3 x analog inputs (scalable), 0 – 5 V, 16 bit resolution
  • 8 x logic level digital control inputs
  • 8 x open collector digital status outputs.

Data Collection

  • 8 GB removable USB flash memory drive that stores the internal data logger, event log, automatic/manual configuration backup and automatic/manual parameter list save
  • Internal data logger (can log up to 12 of 200+parameters)
  • Logger interval (1 second to 24 hours)
  • Storage capacity ~10 years (set to 1 minute interval).

Accessories

Supplied:

Optional:

  • TCP/IP Network Port
  • Dual Sample (particulate) Filter
  • High Pressure Span (for direct cylinder calibration)
  • High Pressure Zero (for direct cylinder calibration)
  • Internal DC Pump (long life, low power, minimal maintenance)
  • Integrated CO2 Sensor (range: 0 – 2000 ppm, accuracy: ± 2 %)
  • Metric Fitting Kit (to connect 6 mm tube to rear ports)
  • 19” Rack Mount Kit
  • External Pump (240 or 110 VAC)
  • Annual Maintenance Kit
  • Printed User Manual (a soft copy of the user manual is supplied on USB with each analyzer).
Specifications are subject to change at any time.